Removal of Polysorbate 80 by complexation prior to LC-MS analysis

Research output: Contribution to journalJournal articleResearchpeer-review

The presence of Polysorbate 80 in samples can challenge liquid chromatography-electrospray ionisation mass spectrometry (LC-ESI-MS) analysis as it is easily ionised and detected. In this study, we demonstrate that interference from Polysorbate 80 can be reduced by complexation with a metal ion followed by precipitation by thiocyanate. The precipitation procedure was tested on a mixture of low molecular weight compounds (e.g. amino acids and non-amino organic acids) and it was shown that none of the tested compounds were precipitated.

Original languageEnglish
JournalAnalusis - European Journal of Analytical Chemistry
Volume408
Issue number9
Pages (from-to)2303-2307
Number of pages5
ISSN0365-4877
DOIs
Publication statusPublished - 2016