High Resolution Mass Spectrometry of Polyfluorinated Polyether-Based Formulation

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

  • Ian Ken Dimzon
  • Trier, Xenia
  • Tobias Frömel
  • Rick Helmus
  • Thomas P. Knepper
  • Pim De Voogt

High resolution mass spectrometry (HRMS) was successfully applied to elucidate the structure of a polyfluorinated polyether (PFPE)-based formulation. The mass spectrum generated from direct injection into the MS was examined by identifying the different repeating units manually and with the aid of an instrument data processor. Highly accurate mass spectral data enabled the calculation of higher-order mass defects. The different plots of MW and the nth-order mass defects (up to n = 3) could aid in assessing the structure of the different repeating units and estimating their absolute and relative number per molecule. The three major repeating units were -C2H4O-, -C2F4O-, and -CF2O-. Tandem MS was used to identify the end groups that appeared to be phosphates, as well as the possible distribution of the repeating units. Reversed-phase HPLC separated of the polymer molecules on the basis of number of nonpolar repeating units. The elucidated structure resembles the structure in the published manufacturer technical data. This analytical approach to the characterization of a PFPE-based formulation can serve as a guide in analyzing not just other PFPE-based formulations but also other fluorinated and non-fluorinated polymers. The information from MS is essential in studying the physico-chemical properties of PFPEs and can help in assessing the risks they pose to the environment and to human health.

OriginalsprogEngelsk
TidsskriftJournal of the American Society for Mass Spectrometry
Vol/bind27
Udgave nummer2
Sider (fra-til)309-318
Antal sider10
ISSN1044-0305
DOI
StatusUdgivet - 2016
Eksternt udgivetJa

Bibliografisk note

Publisher Copyright:
© 2015 The Author(s).

ID: 333814781